Page 1 - Operating Instructions

Operating InstructionsVEGAFLEX 654 … 20 mA/HART - two-wireDocument ID:31845Guided Microwave

Page 2 - Contents

4 Mounting4.1 General instructionsMake sure that all parts of the instrument exposed to the process, inparticular the sensor element, process seal and

Page 3

Fig. 2: Measures against moisture penetrationThe reference plane for the measuring range of the sensors is the sealsurface of the thread.Keep in mind

Page 4 - 1 About this document

In vessels with conical bottom it can b e advantageous to mount thesensor in the center of the vessel, as measurement is then possibledown to the lowe

Page 5 - 2 For your safety

more than 1 m (3.281 in) length, you must secure the probe byfastening a suitable isolated brace or guy directly above the end of therod.VEGAFLEX 65 •

Page 6 - 2.6 CE conformity

5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o

Page 7 - 3 Product description

Caution:No grease should be used when screwing the NPT cable gland orsteel tube into the threaded insert. Standard grease can containadditives that co

Page 8 - 3.3 Operation

2 If an indicating and adjustment module is installed, remove it byturning it to the left.3 Loosen compression nut of the cable entry4 Remove approx.

Page 9

I²CDisplay1 2 5 6 7 83412Fig. 6: Electronics and connection compartment, single chamber housing1 Plug connector for VEGACONNECT (I²C interface)2 Sprin

Page 10 - 4 Mounting

132Display1 2 5 6 7 8I2CFig. 8: Electronics compartment, double chamber housing1 Plug connector for VEGACONNECT (I²C interface)2 Internal connection c

Page 11 - 31845-EN-120329

I2C11 2Fig. 10: Wiring plan, double chamber housing1 Voltage supply, signal output5.5 Wiring plan, Ex-d double chamber housing132Display1 2 5 6 7 8I2C

Page 12

Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .

Page 13

121 2Fig. 12: Connection compartment, Ex-d double chamber housing1 Spring-loaded terminals for power supply and cable screen2 Ground terminal for conn

Page 14 - 5 Connecting to power supply

5.6 Wiring plan - version IP 66/IP 68, 1 bar12Fig. 14: Wire assignment connection cable1 brown (+) and blue (-) to power supply or to the processing s

Page 15 - 5.2 Connection procedure

6 Set up with the indicating and adjustmentmodule PLICSCOM6.1 Short descriptionThe indicating and adjustment module is used for measured valuedisplay,

Page 16

Fig. 15: Insert indicating and adjustment moduleNote:If you intend to retrofit the instrument with an indicating and adjustmentmodule for continuous me

Page 17

6.3 Adjustment system2311.1Fig. 16: Indicating and adjustment elements1 LC display2 Indication of the menu item number3 Adjustment keysl [OK] key:- Mo

Page 18

6.4 Setup stepsAfter connecting VEGAFLEX 65 to power supply or after a voltagerecurrence, the instrument carries out a self-check for approx. 30second

Page 19

If you want to measure the total height of both liquids reliably, pleasecontact our service department or use an instrument speciallydesigned for inte

Page 20

Each product has different reflective properties. In addition, there arevarious interfering factors which have to be taken into account:agitated product

Page 21

Linearisation curveLinearEnter the requested parameters via the a ppropriate keys, save yoursettings and jump to the next menu item with the [->] k

Page 22 - 6.1 Short description

l Unit of measurementl Languagel SensitivityThe following safety-relevant data are not read out or written:l HART model PINl SILl Sensor length/Sensor

Page 23

8 Maintenance and fault rectification8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .378.2 Remove interferences. . . . . . .

Page 24 - 6.3 Adjustment system

Menu item Reset valueLighting No resetLanguage No resetSIL No resetHART mode No resetFactory settingLike basic adjustment, but in addition, special pa

Page 25 - HART mode

6.5 Menu schematicBasic adjustment1▶Basic adjustmentDisplayDiagnosticsServiceInfo1.1Min. adjustment0.00 %=10.000 m(d)8.000 m(d)1.2Max. adjustment100.0

Page 26

Service4Basic adjustmentDisplayDiagnostics▶ServiceInfo4.2ApplicationLiquid▼Standard(DK ≥ 2)▼4.3False signal suppressionChange now?4.4Current outputOut

Page 27

6.6 Saving the parameter adjustment dataWe recommended noting the adjusted data, e.g. in this operatinginstructions manual, and archiving them afterwa

Page 28 - Sensor-TAG

7 Set up with PACTware and other adjustmentprograms7.1 Connect the PC312Fig. 17: Connection of the PC via VEGACONNECT directly to the sensor1 USB cabl

Page 29 - Copy sensor data?

Necessary components:l VEGAFLEX 65l PC with PACTware and suitable VEGA DTMl VEGACONNECTl Power supply unit or processing system12 43OPENTWISTUSBLOCKFi

Page 30

All currently available VEGA DTMs are included as a DTM Collectionon a CD. They can be purchased for a token fee from the responsibleVEGA agency. In a

Page 31 - 6.5 Menu schematic

8 Maintenance and fault rectification8.1 MaintenanceIf the instrument is used properly, no special maintenance is requiredin normal operation.8.2 Remov

Page 32

Error Cause RectificationOperating voltagetoo low or loadresistance toohighCheck, adapt if necessaryCurrent signalgreaterthan 22 mAor less than3.6 mAEl

Page 33

In Ex applications only one instrument and one electronics module withrespective Ex approval may be used.If there is no electronics module available o

Page 34 - 7.1 Connect the PC

1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as

Page 35

At "www.vega.com/downloads" go to "Software". Select under "plicssensors and instruments", "Firmware updates"

Page 36

9 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper

Page 37 - 8.2 Remove interferences

10 Supplement10.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting 316L and TFM PCTFE +2

Page 38

Input variableMeasured variable Level of liquidsMin. dielectric figure of the medium εr> 1.42453124531Fig. 20: Measuring ranges of VEGAFLEX 651 Refe

Page 39 - 8.4 Software update

- 2. HART value (Secondary Value) Distance to the level - scaled (for example hl, %)Resolution, digital > 1 mm (0.039 in)Accuracy (according to DIN

Page 40 - 8.5 Instrument repair

L0-2 mm (-0.079")2 mm (0.079")10 mm (0.394")5 mm (0.197")-5 mm (-0.197")0,15 m (5.906")0,1 m (3.937")0,15 m(5.906&q

Page 41 - 9 Dismounting

0°C(32°F)40°C(104°F)20°C(58°F)60°C(140°F)80°C(176°F)50°C(122°F)-40°C(-40°F)100°C(212°F)150°C(302°F)12Fig. 23: Ambient temperature -

Page 42 - 10 Supplement

Spring-loaded terminals for wire cross-sec-tion< 2.5 mm² (AWG 14)Electromechanical data - version IP 66/IP 68 (1 bar)Cable entry- Single chamber ho

Page 43

Permissible residual ripple- < 100 Hz Uss< 1 V- 100 Hz … 10 kHz Uss< 10 mVLoad see diagram1000750500250141816 20 22 24 26 28 30 32 34 36ΩV412

Page 44

- double channel diversitary redundantarchitecture (1oo2D)up to SIL3You can find detailed information in the supplied Safety Manual of the instrument s

Page 45

2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist

Page 46

10.2 DimensionsThe following dimensional drawings represent only an extract of thepossible versions. Detailed dimensional drawings can be downloadedon

Page 47

Aluminium housing in protection rating IP 66/IP 68, 1 bar120 mm (4.72")~ 105 mm (4.13")116 mm (4.57")~ 150 mm (5.91")ø 84 mm (3.31

Page 48

VEGAFLEX 65, threaded versionL 36 (G¾, ¾ NPT) 41 (G1, 1 NPT) 46 (G1½, 1½ NPT) G¾, ¾ NPT, G1, 1 NPT, G1½, 1½ NPT 46 mm(1 13/16")27mm (1 1/16"

Page 49

VEGAFLEX 65, flange versionLbLbDd dDkk68mm(2.68")46mm(1.81")D b kdDN 25 PN 40 115 18 85 4x ø 14DN 40 PN 40 150 18 110 4x ø 181" 150 lb

Page 50 - 10.2 Dimensions

10.3 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only

Page 51 - Stainless steel housing

VEGAFLEX 65 • 4 … 20 mA/HART - two-wire 5510 Supplement31845-EN-120329

Page 52 - , threaded version

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9

Page 53 - , flange version

2.5 Safety label on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fulfills the

Page 54 - 10.4 Trademark

3 Product description3.1 StructureThe scope of delivery encompasses:l Level sensor VEGAFLEX 65l Documentation- this operating instructions manual- Saf

Page 55

The type label contains the most important data for identification anduse of the instrument:l Article numberl Serial numberl Technical datal Article nu

Page 56 - ISO 9001

l With a HART handheld3.4 Packaging, transport and storageYour instrument was protected by packaging during transport. Itscapacity to handle normal lo