Operating InstructionsVEGAWAVE 61- two-wireDocument ID:32250Vibration
4 Mounting4.1 General instructionsMake sure that all parts of the instrument exposed to the process, inparticular the sensor element, process seal and
The max. permissible pressure is specified in chapter "Technical data"or on the type label of the sensor.The vibrating level switch is a meas
To achieve a very precise switching point, you can install VEGAWAVE61 horizontally. However, if the switching point can have a tolerance ofa few centi
12Fig. 5: Flow orientation of the tuning fork1 Marking with screwed version2 Direction of flowIn case of horizontal mounting in adhesive products, the
> 125 mm(> 5")> 120 mm(> 4.7")Fig. 7: Baffle for protection against mechanical damage14 VEGAWAVE 61 • - two-wire4 Mounting32250-EN
5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o
5 Lift the opening levers of the terminals with a screwdriver (seefollowing illustration)Fig. 8: Connection steps 5 and 66 Insert the wire ends into t
14442 3Fig. 9: Material versions, single chamber housing1 Plastic (not with EEx d)2 Aluminium3 Stainless steel (not with EEx d)4 Filter element for pr
5.4 Wiring plan - version IP 66/IP 68, 1 bar12Fig. 11: Wire assignment connection cable1 brown (+) and blue (-) to power supply or to the processing s
6 Set up6.1 General informationThe figures in brackets refer to the following illustrations.On the electronics module you will find the following indica
Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .
6.2 Adjustment elements34125Fig. 12: Electronics and connection compartment - two-wire output1 Potentiometer for switching point adaptation2 DIL switc
When used on a control system, the following values apply:l Mode min.- Vibrating element uncovered - 16 mA ±1 mA- Vibrating element covered - 8 mA ±1
VEGAWAVE 61 level switch with signal conditioning instrument1)The following chart provides an overview of the switching conditionsdepending on the adj
This is possible up to SIL2 in a single channel architecture (1oo1D),and up to SIL3 in a multiple channel, redundant architecture.The following instru
The correctness of the subsequent switching conditions must bemonitored via the two LEDs on the signal conditioning instrument aswell as the connected
signal conditioning instrument). Keep in mind that connected instru-ments are activated during the function test. By doing this, you cancheck the corr
Test not passedl Interference signal (< 3.6 mA) < 400 ms / ≥ 750 msl Uncovered (approx. 8 mA) < 1 s / ≥ 2 sl Covered (approx. 16 mA) < 1 s
7 Maintenance and fault rectification7.1 MaintenanceIf the instrument is used properly, no special maintenance is requiredin normal operation.7.2 Remov
Error Cause RectificationWrong mode selected Set the mode switch on VEGAWAVE 61 to "max". Setthe correct mode on the signal conditioning inst
6 Pull out the old electronics module7 Compare the new electronics module with the old one. The typelabel of the electronics module must correspond to
9 Supplement9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .319.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . .
8 Dismounting8.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper
9 Supplement9.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting - thread 316L- Process
Output variableOutput Two-wire outputSuitable signal conditioning instruments VEGATOR 536Ex, 537Ex, 636ExOutput signal- Mode min. Vibrating element un
Product density- Standard > 0.02 g/cm³ (0.0007 lbs/in³)- adjustable > 0.008 g/cm³ (0.0003 lbs/in³)Granular size max. 10 mm (0.4 in)Electromechan
Voltage supplyOperating voltage 10 … 36 V DC (via the signal conditioning instru-ment)Electrical protective measuresProtection rating- Plastic housing
9.2 DimensionsHousing in protection IP 66/IP 67 and IP 66/IP 68; 0.2 bar3214~ 69 mm(2.72")ø 79 mm(3.11")117 mm (4.61")M20x1,5/½ NPT~ 59
22mm(55/64")33mm(1 19/64")220mm (8 21/32")150mm (5 29/32")G1½Aø 43mm (1 11/16")Fig. 52: VEGAWAVE 61, threaded version G1½ A (
ø 34 mm(1 11/32")178 mm (7 1/64")Fig. 53: Temperature adapterVEGAWAVE 61 • - two-wire 379 Supplement32250-EN-120418
9.3 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only i
VEGAWAVE 61 • - two-wire 399 Supplement32250-EN-120418
1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as
VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9
2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist
2.5 Safety label on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 Safety label on the instrumentThe saf
3 Product description3.1 StructureThe scope of delivery encompasses:l VEGAWAVE 61 point level switchl Documentation- this operating instructions manua
3.2 Principle of operationVEGAWAVE 61 is a point level sensor with tuning fork for leveldetection.It is designed for industrial use in all areas of pr
l Potentiometer for adaptation to the product densityl Mode changeover for selection of the output current3.4 Storage and transportYour instrument was
Commentaires sur ces manuels