Operating InstructionsRadar sensor for continuous level measurement of liquidsVEGAPULS 65Modbus and Levelmaster protocolDocument ID: 41365
103 Product descriptionVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210113 Product description3.1 CongurationThe type label contains the
113 Product descriptionVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101121Fig. 5: Position of Modbus and sensor electronics1 Modbus electr
123 Product descriptionVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210113.3 Packaging, transport and storageYour instrument was protected
133 Product descriptionVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Screwed anges are available in dierent versions according to the
144 MountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210114 Mounting4.1 General instructionsWith instruments with threaded process tti
154 MountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210111Fig. 6: Position of the polarisation level1 Marking holeWhen mounting the V
164 MountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Fig. 8: Mounting of the radar sensor on vessels with conical bottomDo not moun
174 MountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Fig. 10: Recommended socket mountingIn case of very big socket diameters, the
184 MountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Fig.12:CoversmoothproleswithdeectorsIf there are agitators in the vess
195 Connecting to power supply and bus systemVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210115 Connecting to power supply and bus system5
2Quick startVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Quick startThe quick start enables a quick setup with many applications. You c
205 Connecting to power supply and bus systemVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The connection to the indicating and adjustme
215 Connecting to power supply and bus systemVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101110. Place probably existing indicating and a
225 Connecting to power supply and bus systemVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011++power supplyMODBUSD0D1IS GNDUSB1213452off
236 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210116 Set up the sensor with t
246 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210116.3 Adjustment system12Fig
256 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Setup: Settings, for exampl
266 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The following options are a
276 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• Product speed: – Fast to
286 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• Process/measurement condi
296 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011If you want to measure the
3Quick startVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210113. Select in the menu item "Application" the vessel, the applicatio
306 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• Adjustment for all applic
316 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011100%0%211 m(39.37")10
326 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210114. Enter the suitable dist
336 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The "Echocurve"
346 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• Vessel installations such
356 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The menu item "Delete&
366 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• Setup: Resetting of the p
376 Set up the sensor with the indicating and adjustment moduleVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011If the instrument is equipp
387 Setting up sensor and Modbus interface with PACTwareVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210117 Setting up sensor and Modbus in
397 Setting up sensor and Modbus interface with PACTwareVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The connection of the PC to the RS
4ContentsVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Contents1 About this document1.1 Function ...
407 Setting up sensor and Modbus interface with PACTwareVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Fig.24:ExampleofaDTMviewAll
418 Set up with other systemsVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210118 Set up with other systems8.1 DD adjustment programsDevice
429 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210119 Diagnosis, Asset Management and service9.1 M
439 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Further echo curves: Up to 10 echo curves can b
449 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The following table shows the codes and text me
459 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011CodeText mes-sageCause Rectication PA DevS-pec
469 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011The following table shows the error codes and t
479 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Depending on the reason for the fault and the m
489 Diagnosis, Asset Management and serviceVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011You can nd detailed information on www.vega.co
4910 DismountingVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101110 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be aware
5ContentsVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210119.4 Rectify faults ...
5011 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111 Supplement11.1 Technical dataGeneral data316L corresponds to 1.4404 or
5111 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210113421Fig. 26: Data of the input variable1 Reference plane2 Measured variab
5211 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210111,0 m (3.28 ft)20 mm (0.788 in)8 mm (0.315 in)- 8 mm (- 0.315 in)- 20 mm (-
5311 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Process conditionsFor the process conditions, please also note the specica
5411 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Integrated clockDate format Day.Month.YearTime format 12 h/24 hTime zone Ex
5511 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011VEGAPULS 65 and is activated/deactivated via a slide switch.413 3 32++power
5611 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.2 GeneralcongurationofthehostThe data exchange with status and var
5711 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Register Name Register Number Type NoteSV 110 Secondary Variable in Byte Or
5811 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Register Name Register Number Type NotePV 2202 DWord Primary Variable in By
5911 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.4 FC4 Read Input RegisterWith this command, any number (1-127) of input
61 About this documentVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210111 About this document1.1 FunctionThis operating instructions manual
6011 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Implemented function codes:Sub Function Code Name0x00 Return Data Request0x
6111 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Parameter Length Code/DataMEI Type 1 Byte 0x0ERead Device ID Code 1 Byte 0x
6211 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011• "*" can be used as a joker for any position in the address. The
6311 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.5 Set number of FloatsRequest:Parameter Length Code/DataSet number of F
6411 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Response:Parameter Length Code/DataSet Receive to Transmit Delay 5 characte
6511 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.6 CongurationoftypicalModbushosts11.6 Fisher ROC 809Wiring plan++
6611 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.6 ABB Total FlowWiring plan++power supplyMODBUSD0D1IS GNDUSB1 213452o
6711 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.6 Thermo Electron AutopilotWiring plan++power supplyMODBUSD0D1IS GNDU
6811 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.6 Bristol ControlWave MicroWiring plan++power supplyMODBUSD0D1IS GNDU
6911 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.6 ScadaPackWiring plan++power supplyMODBUSD0D1IS GNDUSB132134552134967
72 For your safetyVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210112 For your safety2.1 Authorised personnelAll operations described in t
7011 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011Housing1~ 87 mm (3.43")M16x1,5ø 84 mm(3.31")120 mm (4.72")M2
7111 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011VEGAPULS 65, threaded version41 mm(1.61")ø 40 mm(1.58")ø 33 mm(1.
7211 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011VEGAPULS65,angeversionø 33 mm(1.30")395 mm (15.55")120 mm (4.
7311 SupplementVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-12101111.8 Industrial property rightsVEGA product lines are global protected by
74INDEXVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011INDEXAAccessories – Flanges 13 – Indicating and adjustment module 12 – Interface
75NotesVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011
Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany41365-EN-121011All statements concerning scope of delivery, application, practi
82 For your safetyVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-1210112.5 CE conformityThe device fullls the legal requirements of the appli
92 For your safetyVEGAPULS 65 • Modbus and Levelmaster protocol41365-EN-121011tection. The environment management system is certied according to DIN
Commentaires sur ces manuels