Level and Pressure
Operating Instruction
VEGAPULS 56V
Vue de la page 0
1 2 3 4 5 6 ... 79 80

Résumé du contenu

Page 1 - Operating Instruction

Level and PressureOperating InstructionVEGAPULS 56V

Page 2 - Note Ex-area

10 VEGAPULS 56V2 Types and versionsVEGAPULS 56 sensors are a new developedgeneration of very compact high temperatureradar sensors. For the first time

Page 3 - VEGAPULS 56V 3

VEGAPULS 56V 11VEGAPULS 56 V EXXX X X X X X X XJ - Tube extension for horn antennaX - withoutA - Aluminium housingD - Aluminium housing with Exd-conne

Page 4 - 1 Product description

12 VEGAPULS 56V2.2 Configuration of measuringsystemsA measuring system consists of a sensorand a processing unit. The processing unit(the signal condi

Page 5 - Pulse sequence

VEGAPULS 56V 132244VEGAMET515V2Measuring system with 1 … 2 VEGAPULS 56V on signal conditioning instrumentVEGAMET 515V• Two-wire technology, supply fro

Page 6 - 1.2 Application features

14 VEGAPULS 56V2244VEGA MET515VVEGATR ENN5472Types and versionsMeasuring system with 1 … 2 VEGAPULS 56V Ex, 56V Ex0 via a separatorVEGATRENN 548V Ex o

Page 7 - 1.3 Adjustment

VEGAPULS 56V 152244VEGAMET515V2Measuring system with 1 … 2 VEGAPULS 56V Ex, 56V Ex0 with pressuretight encapsulated connection box on the signal condi

Page 8 - 8 VEGAPULS 56V

16 VEGAPULS 56VVEGA LOG571 CPUVEGA LOG571 EV22CPU22242424Measuring system with 15 VEGAPULS 56V via one two-wire line on the processingsystem VEGALOG

Page 9 - 1.4 Antennas

VEGAPULS 56V 17Measuring system with five VEGAPULS 56V Ex, 56V Ex0 per two-wire linevia separator VEGATRENN 548V Ex on the processing systemVEGALOG 57

Page 10 - 2 Types and versions

18 VEGAPULS 56VVEGAL OG571 CPUVEGAL OG571 EV22CPU22222444Measuring system with 15 VEGAPULS 56V Ex, 56V Ex0 with pressure tightencapsulated connectio

Page 11 - Type code

VEGAPULS 56V 193 Technical data3.1 DataPower supplySupply voltage from signal conditioning instrument VEGAMETor processing system VEGALOG 571(max. 36

Page 12 - 12 VEGAPULS 56V

2 VEGAPULS 56VContentsSafety informationThe described module must only be installedand operated as described in this operatinginstruction. Please note

Page 13 - VEGAMET 515V

20 VEGAPULS 56VPN 16PN 25PN 40PN 64bar402516-40 0 50 100 150 200 250 300 350˚C64Flange DIN DN 50Material: 1.4571Groove and tongue acc.to DIN 2512 form

Page 14 - VEGATR ENN

VEGAPULS 56V 21Technical databar402516-40 0 50 100 150 200 250 300 350˚CPN 16PN 25PN 40PN 16PN 25PN 40PN 64bar402516-40 0 50 100 150 200 250 300 350˚C

Page 15

22 VEGAPULS 56VFlange DIN DN 200Material: 1.4571Groove and tongue acc.to DIN 2512 form F, NPN 16PN 25PN 40PN 64bar40251664-40 0 50 100 150 200 250 300

Page 16 - 571 CPU

VEGAPULS 56V 23Technical dataFlange DIN DN 250Material: 1.4571Groove and tongue acc.to DIN 2512 form F, NPN 16PN 25PN 40PN 64bar40251664-40 0 50 100 1

Page 17 - VEGALOG 571

24 VEGAPULS 56VEx-technical data (note the approval documents in the yellow binder)Classification- d pressure tight encapsulation- ia intrinsically sa

Page 18 - VEGAL OG

VEGAPULS 56V 25Weights in kg (1 psi = 0,0689 bar)DIN 16 bar 25 bar 40 bar 64 bar- DN 50 6,9 -- 7,7 8,5- DN 80 8,8 -- 10,0 10,9- DN 100 9,8 -- 11,7 14,

Page 19 - 3 Technical data

26 VEGAPULS 56V20120ø180ø96ø220ø18ø160ø7675ø200ø18ø125ø1820ø1651823DN 50DN 80DN 10032021318537020511625185213320370205116253.2 DimensionsTechnical

Page 20 - 20 VEGAPULS 56V

VEGAPULS 56V 27Technical data22205ø146ø285DN 150ø240 ø2226380ø241ø355ø26DN 250ø40524296ø197ø295ø22DN 200ø340

Page 21

28 VEGAPULS 56VFlange dimensions according to ANSID = outer flange diameterb = flange thicknessk = diameter of hole circled1= seal ledge diameterf = s

Page 22

VEGAPULS 56V 293.3 ApprovalsWhen using radar sensors in Ex and StEx-areas or on ships the instruments must besuitable and approved for these explosion

Page 23

VEGAPULS 56V 3Contents4 Mounting and installation4.1 General installation instructions ... 304.2 Measurement of

Page 24 - 24 VEGAPULS 56V

30 VEGAPULS 56VMeasuring range (operating range) and max. measuring distanceNote: The use of the sensors in solid applications is restricted.fullMeas.

Page 25 - Outputs and processings

VEGAPULS 56V 31010 m0m20 m20••30••50 %3,5 m5,0 m3,5 m5,0 m100 %010 m0m20 m30••40••100 %50 %5,3 m5,3 m6,8 m 6,8 mEmission cone of a DN!100 horn antenna

Page 26 - 3.2 Dimensions

32 VEGAPULS 56V02,4 m3,8 m2,4 m3,8 m10 m0m20 m14••22••50 %100 %Heat isolationMounting on a longer DIN-socket piece>!10!mmEmission cone of a DN 250

Page 27 - Technical data

VEGAPULS 56V 33Reference planeMounting on dished vessel endReference plane1/2 vessel radiusMounting and installationMounting directly on flat vessel t

Page 28 - Adjustment module MINICOM

34 VEGAPULS 56V0%75 %100 %100 %0%Pipe flange system as bypass pipeExtended bypass pipe on the vessel with strongproduct movementsMounting and installa

Page 29 - 3.3 Approvals

VEGAPULS 56V 35Openings in a surge pipe for mixing of inhomogene-ous productsMounting and installationslightly inhomogeneousliquidshomogeneousliquidsP

Page 30 - 4 Mounting and installation

36 VEGAPULS 56VThe sensor must be directed with the hole to the rowsof holes or openings.HoleDN 50ø50Mounting and installationLockable measuring pipe

Page 31 - VEGAPULS 56V 31

VEGAPULS 56V 37Pipe antenna: The surge pipe open to the bottommust have a ventilation or a compensation hole.CorrectWrongMounting and installationA de

Page 32 - 4.2 Measurement of liquids

38 VEGAPULS 56V0%~45••ø 51,20,0…0,41,5…22,90,0...0,4150…5005…152,9…6100 %Mounting and installationConstruction instructions for standpipeRadar sensors

Page 33 - (surge or bypass pipe)

VEGAPULS 56V 39Mounting and installation0%~45••ø 100,80,0…0,41,5…23,60,0…0,4150…5005…153,6ø 962100 %Flange DN 100Smooth weldingflangeConnectingsleeveW

Page 34 - 34 VEGAPULS 56V

4 VEGAPULS 56VProduct description1 Product descriptionLevel measurement on high temperatureprocesses or on mediums with high tempera-ture was formerly

Page 35 - Adhesive products

40 VEGAPULS 56VMounting and installation4.4 False echoesThe installation place of the radar sensormust be chosen such that no struts orinflowing mater

Page 36 - Standpipe with ball valve

VEGAPULS 56V 4175 %0%100 %Mounting and installationHeavy turbulencesBuild-upInflowing liquidWrongCorrectStrong product movementsHeavy turbulences in t

Page 37 - Ventilation holes

42 VEGAPULS 56VDirect sensor vertically to the product surfaceCorrect WrongLadderLadderMounting and installationHorn antenna: Correct and wrong socket

Page 38 - 38 VEGAPULS 56V

VEGAPULS 56V 43Mounting and installationThe polarisation direction is in line with the hole. Thesensor must be directed with the hole to the rows ofho

Page 39 - VEGAPULS 56V 39

44 VEGAPULS 56VElectrical connection5 Electrical connection5.1 Connection and connectioncableSafety informationEnsure that the instrument is in curren

Page 40 - 4.4 False echoes

VEGAPULS 56V 45Electrical connectionFigure 1: Earthing only on the sensor sideFigure 2: Earthing on both sides (on the signal conditioning instrument

Page 41 - Build-up

46 VEGAPULS 56V- +21ShieldSupply: 20 … 36 V DC, VBUS12-+ESC+-OK1 2 C 567843VBUSCommu-nication+-4...20mADisplay12 C 567843EEx d-connection housing(open

Page 42 - >10 mm

VEGAPULS 56V 47SENSORDISPLAY1234 5678ESC+-OK-+1 2 C 567843Commu-nication+-4...20mADisplay12 C 567843VBUS-+ESCOKTank 1m (d)12.345Electrical connection5

Page 43 - VEGAPULS 56V 43

48 VEGAPULS 56VSet-upAdjustment with the adjustment moduleMINICOMWith the adjustment module MINICOM youadjust directly in the sensor or in the externa

Page 44 - 5 Electrical connection

VEGAPULS 56V 49When the signal conditioning instruments areexceptionally not pre-configured, start nowwith the following section "Configuration“

Page 45 - VEGAPULS 56V 45

VEGAPULS 56V 5Time transformationPulse sequenceMeas. distanceemission - reflection - receipt1 ns278 nsttProduct descriptionThe radar impulses are emit

Page 46 - Ex ia-version

50 VEGAPULS 56V• Click in the menu window "Create newmeasurement loop - Measurement loopdesignation“ to "Level“.Set-up• Choose "Standar

Page 47 - DISPLAY1234 5678

VEGAPULS 56V 51In the heading of the opening menu windowyou now see the previously entered mea-surement loop name and the description.• First choose &

Page 48 - 6 Set-up

52 VEGAPULS 56V• Confirm your adjustments with "OK“ andyou are again in the menu window "Adjust-ment“.• Click in the menu window "Adjus

Page 49 - "Parameter adjustment

VEGAPULS 56V 53As parameter you can choose "dimension-less (figures), volume, mass, height anddistance“ and an appropriate unit (e.g. l. hl)can t

Page 50 - 50 VEGAPULS 56V

54 VEGAPULS 56VFirst of all a linear relation (a straight line) isdisplayed. In the field "Transfer measuredvalue“ the present level in percent o

Page 51 - Parameter adjustment

VEGAPULS 56V 55Again in the menu window "Conditioning“ youcan enter with the menu point "Integrationtime“ a measured value integration. This

Page 52 - 52 VEGAPULS 56V

56 VEGAPULS 56V• Choose under "Scaling for sensor display“e.g. liter.• Enter the measuring distance in meter,which you have entered in the min./m

Page 53 - VEGAPULS 56V 53

VEGAPULS 56V 57If a smaller measuring range is selected aspreviously adjusted in the "Min/Max-adjust-ment“, the sensor is caused to provide e.g.t

Page 54 - Linearisation

58 VEGAPULS 56VEcho curveWith the menu point "Echo curve“ you candisplay the course and the strength of thedetected radar echo.If you have to exp

Page 55 - VEGAPULS 56V 55

VEGAPULS 56V 59• Enter here the checked product distanceand click to "Create new“.Hence you cause the sensor to mark all ech-oes in front of the

Page 56 - Sensor optimisation

6 VEGAPULS 56V1.2 Application featuresApplications• level measurement of liquids and solids• measurement also in vacuum• all slightly conductive mater

Page 57 - Meas. environment

60 VEGAPULS 56V• Close the information windows.• Quit the menu "Measurement loop data“.• Click in the menu window "Instrument dataparameter

Page 58 - 58 VEGAPULS 56V

VEGAPULS 56V 61Indicating and adjustment surfaces of:ESCOK-+12on100%CONNECT515 VDependent on the menu point,change the value or choose out ofthe listC

Page 59 - VEGAPULS 56V 59

62 VEGAPULS 56VSet-upThe most important adjustment stepsOn page 66␣ …␣ 73 you see the completemenu plan of the signal conditioning instru-ment VEGAMET

Page 60 - 60 VEGAPULS 56V

VEGAPULS 56V 63Max.Min.100 % (0,500 m) corresp. to1200 litersMeas. range0 % (3,400 m) corresp. to45!litersSet-upAdjust-mentPa-ram-eteradjust-mentSenso

Page 61 - VEGAPULS 56V 61

64 VEGAPULS 56VSignalconditioningSca-ling0 %corre-spondsXXXX100 %corre-spondsXXXXDeci-malpoint888.8prop.toMassUnitKgAdjustment with mediumExample:Min/

Page 62 - 62 VEGAPULS 56V

VEGAPULS 56V 65Ampl.:XXdBS-N:XX dBLin.curveLinearIntegrationtime0 sAdjust-mentSignalconditioningSca-lingLinearisation:Pre-adjusted is a linear depende

Page 63 - Adjustment

66 VEGAPULS 56VParam.TAG-No. 1TAG -No. 1 % xx,xAdjustmentSignalconditioningOut-putsmAout-putsSca-lingLIN-curvelinearmAoutputno. 1Voltoutputno. 1p

Page 64 - 5) Operating range

VEGAPULS 56V 67MET-dis-playPC/DCSout-putsDIS-out-putsSimu-lationSpe-cialfunct.ResetlevelResetOK ?ResetNow!OK ?Simu-lation % XX,XSimu-lationNow! OK?D

Page 65 - Conditioning

68 VEGAPULS 56VConfiginputsConfigmeas.loopInputno. 1Appli-cationlevelSensortypeHydro-staticModeStan-dardOption nooptionSensorcoordinationLoca-tion AIn

Page 66 - 66 VEGAPULS 56V

VEGAPULS 56V 69Configout-putsConfigcurr.output.mA no1toTAG -No. 1V no1toTAG -No. 1ConfigVolt.outputConfigrelayoutputConfigPC/DCSoutputConfigVEGA-DISDI

Page 67 - VEGAPULS 56V 67

VEGAPULS 56V 7One or two sensors on the signal conditioning instru-ment; adjustment with the PC on the signal condition-ing instrumentWith the standar

Page 68 - 68 VEGAPULS 56V

70 VEGAPULS 56VAdd’lfunc-tionsPass-wordoffEditlin.curvesInfoLang-uageEng-lishInputinfoVEGA-METInfoPro-graminfoMeas.loopinfoPro-graminfoxxxxxxTAG 1TAG

Page 69 - VEGAPULS 56V 69

VEGAPULS 56V 71ResetVEGA-METResetconfi-gura-tionResetTAG 1to de-faultReset OK?ResetNow! OK?tolinkedappli-cationDeleteallTAG’s?Reset OK?ResetNow! O

Page 70 - 70 VEGAPULS 56V

72 VEGAPULS 56VSensoropti-mizeSensorTagSensorMeas.environmentOpera-tingrangeBeginm (d) 0.50Endm (d)6.00Meas.ConditionsConditionliquidFastchangeNoAgita

Page 71 - VEGAPULS 56V 71

VEGAPULS 56V 73ResettodefaultDis-tancem (d)4.700ResetNow ! OK ?ResetrunsAdd’lfunc-tionsInfoLan-guageEng-lishResetNow !OK ?Resetruns!ResettodefaultMea

Page 72 - MINICOM - Menu plan

74 VEGAPULS 56V• Click to "Yes“.When you connect VVO to a VEGALOG onwhich data had been already saved, you geta message if you want transfer the

Page 73 - VEGAPULS 56V 73

VEGAPULS 56V 75• Click in the card-info-window of your EV-card (input VBUS) to "Sensor survey“.You get a survey of the sensors connected tothe ca

Page 74 - User access

76 VEGAPULS 56V• Choose in the next menu window "Stand-ard level measurement“ and "no options“.• Click to "Continue“.After a few second

Page 75 - Configuration measuring

VEGAPULS 56V 77loop - Measurement loop designation“ opens.Enter here:– a measurement loop number– a measurement loop description– and coordinate to yo

Page 76 - 76 VEGAPULS 56V

78 VEGAPULS 56VParameter adjustment (adjustment,conditioning, linearisation, gauging byliters, outputs)Sensor optimisation (meas. environ-ment, echo c

Page 77 - VEGAPULS 56V 77

VEGAPULS 56V 79Notes

Page 78 - Configuration/Measuring sys

8 VEGAPULS 56VProduct descriptionAdjustment with adjustment moduleMINICOMWith the (3,2␣ cm␣ x␣ 6,7␣ cm) 6-key adjustmentmodule with display you can ca

Page 79 - VEGAPULS 56V 79

ISO 9001The statements on types, application, use and operating conditions ofthe sensors and processing systems correspond to the actualknowledge at t

Page 80 - ISO 9001

VEGAPULS 56V 9Pipe antennaPipe antennas on surge or by-pass pipes only form a completeantenna system in conjunctionwith a measuring pipe which canalso