Operating InstructionsVEGAWAVE 61- NAMURDocument ID:32249Vibration
4 Mounting4.1 General instructionsMake sure that all parts of the instrument exposed to the process, inparticular the sensor element, process seal and
The max. permissible pressure is specified in chapter "Technical data"or on the type label of the sensor.The vibrating level switch is a meas
To achieve a very precise switching point, you can install VEGAWAVE61 horizontally. However, if the switching point can have a tolerance ofa few centi
12Fig. 5: Flow orientation of the tuning fork1 Marking with screwed version2 Direction of flowIn case of horizontal mounting in adhesive products, the
> 125 mm(> 5")> 120 mm(> 4.7")Fig. 7: Baffle for protection against mechanical damage14 VEGAWAVE 61 • - NAMUR4 Mounting32249-EN-12
5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o
5 Lift the opening levers of the terminals with a screwdriver (seefollowing illustration)Fig. 8: Connection steps 5 and 66 Insert the wire ends into t
14442 3Fig. 9: Material versions, single chamber housing1 Plastic (not with EEx d)2 Aluminium3 Stainless steel (not with EEx d)4 Filter element for pr
+-+-1 2 3 4321Fig. 11: Wiring plan - External simulation key1 NAMUR amplifier2 Bridge3 External simulation key5.4 Wiring plan - version IP 66/IP 68, 1
6 Set up6.1 General informationThe figures in brackets refer to the following illustrations.On the electronics module you will find the following indica
Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .
By default, the potentiometer of VEGAWAVE 61 is set to the right stop(> 0.02 g/cm³ or 0.0008 lbs/in³). In case of very light-weight solids, turnthe
Level Signal current -SensorControl lampFalling characteris-tics max.≥ 2.4 mAFalling characteris-tics max.≤ 1 mARising characteris-tics min.≥ 2.4 mARi
You can carry out the function test with the outputted current valuesalso directly via a safety PLC or a process control system.A function test can be
Sensor cur-rentLevel relayamplifier -overflowprotectionLevel relayamplifier -dry run pro-tectionSignal lampamplifier -overflowprotectionSignal lampamplifier
7 Maintenance and fault rectification7.1 MaintenanceIf the instrument is used properly, no special maintenance is requiredin normal operation.7.2 Remov
Error Cause RectificationWrong characteristic se-lectedSet the correct characteristics on the characteristicsreversal switch (overflow protection, dry r
21Fig. 30: Loosening the holding screws1 Electronics module2 Screws (2 pcs.)6 Pull out the old electronics module7 Compare the new electronics module
You can download a return form (23 KB) from our homepage at www.vega.com under: "Downloads - Forms and certificates - Repair form".By doing t
8 Dismounting8.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper
9 Supplement9.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting - thread 316L- Process
9 Supplement9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . .
Output variableOutput Two-wire NAMUR outputCurrent consumption- Falling characteristics (max.) ≥ 2.4 mA uncovered/≤ 1 mA covered- Rising characteristi
Product density- Standard > 0.02 g/cm³ (0.0007 lbs/in³)- adjustable > 0.008 g/cm³ (0.0003 lbs/in³)Granular size max. 10 mm (0.4 in)Electromechan
Simulation key simulation of a line break between sensor andprocessing unitVoltage supplyOperating voltage (characteristics accordingto standard)for c
9.2 DimensionsHousing in protection IP 66/IP 67 and IP 66/IP 68; 0.2 bar3214~ 69 mm(2.72")ø 79 mm(3.11")117 mm (4.61")M20x1,5/½ NPT~ 59
22mm(55/64")33mm(1 19/64")220mm (8 21/32")150mm (5 29/32")G1½Aø 43mm (1 11/16")Fig. 35: VEGAWAVE 61, threaded version G1½ A (
ø 34 mm(1 11/32")178 mm (7 1/64")Fig. 36: Temperature adapterVEGAWAVE 61 • - NAMUR 359 Supplement32249-EN-120418
9.3 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only i
VEGAWAVE 61 • - NAMUR 379 Supplement32249-EN-120418
38 VEGAWAVE 61 • - NAMUR9 Supplement32249-EN-120418
VEGAWAVE 61 • - NAMUR 399 Supplement32249-EN-120418
1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as
VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9
2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist
2.5 Safety label on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fulfills the
3 Product description3.1 StructureThe scope of delivery encompasses:l VEGAWAVE 61 point level switchl Documentation- this operating instructions manua
3.2 Principle of operationVEGAWAVE 61 is a point level sensor with tuning fork for leveldetection.It is designed for industrial use in all areas of pr
l Potentiometer for adaptation to the product densityl Mode switch to select the switching condition (reverse character-istics)l Simulation key3.4 Sto