Operating Instructions
Lock tting ARV-VB63.B
for VEGAVIB 63
pressure range -1 25 bar
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1 - Operating Instructions

Operating InstructionsLock fitting ARV-VB63.Bfor VEGAVIB 63– pressure range -1 … 25 bar

Page 2 - Contents

10 Screw the pressure screw (10) with a view turns from topinto the mounting boss (6)11 Clean the connection tube of the sensor and the surfacesof the

Page 3 - 1 About this document

6 Dismounting6.1 Dismounting procedureNote chapter "Mounting" and carry out the described steps inreverse order.Warning:If you want to remov

Page 4 - 2 For your safety

7 Supplement7.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Process fitting- G1½A or 1½ NPTTube diameter of the sensor ø 29

Page 5 - 3 Product description

7.2 DimensionsLock fitting ARV-VB63.B for VEGAVIB 63~ 67mm (2 41/64")20mm(25/32")55mm(2 11/64")ø 29mm(1 9/64")G2A/2NPTG1½A/1½NPT55m

Page 6 - 4 Mounting

14 Lock fitting ARV-VB63.B -–pressure range -1 … 25 bar20036-EN-050804

Page 7

Lock fitting ARV-VB63.B -–pressure range -1 … 25 bar 1520036-EN-050804

Page 8

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] 9001All stateme

Page 9 - 5.2 Exchange seals

Contents1 About this document1.1 Function ...41.2 Target group ...41.3 Symbolism used ...

Page 10 - 5.3 Instrument repair

1 About this document1.1 FunctionThis operating instructions manual has all the information youneed for quick setup and safe operation of ARV-VB63.B.P

Page 11 - 6 Dismounting

2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manualmust be carried out only by trained, specialise

Page 12 - 7 Supplement

3 Product description3.1 ConfigurationThe scope of delivery encompasses:l Lock fitting ARV-VB63.B for VEGAVIB 63 vibrating levelswitchl Documentation- t

Page 13 - 7.2 Dimensions

4 Mounting4.1 General instructionsThe note of the following general safety instructions:l Dismount the lock fitting only in unpressurized conditionl Us

Page 14 - 20036-EN-050804

7 Tighten the terminal screws (2) alternately. Make sure thatthe gap between clamps and pressure screw is in parallel.Tighten the terminal screws up t

Page 15

12453789768101111Fig. 1: Lock fitting ARV-VB63.B - up to 25 bar (362 psi)1 Spring ring B62 Terminal screw M6 x 253 Locking angle4 Spring ring B45 Holdi

Page 16 - ISO 9001

5 Maintenance and fault rectification5.1 MaintenanceWhen used as directed in normal operation, lock fitting ARV-VB63.B is completely maintenance-free.5.