Résumé du contenu

Page 1 - VEGAPULS 62 standpipe ver

Operating InstructionsRadar sensor for continuous level measurement of liquidsVEGAPULS 62 standpipe ver-sionFoundation FieldbusDocument ID: 36510

Page 2 - Quick start

103 Product descriptionVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210113 Product description3.1 CongurationThe type label contain

Page 3

113 Product descriptionVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The instrument is available in two dierent electronics versi

Page 4 - Contents

123 Product descriptionVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210113.3 Packaging, transport and storageYour instrument was prot

Page 5

133 Product descriptionVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011It is suitable for measured value indication and adjustment o

Page 6 - 1 About this document

144 MountingVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210114 Mounting4.1 General instructionsWith instruments with threaded proces

Page 7 - 2 For your safety

155 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210115 Connecting to the bus system5.1 Preparing the co

Page 8 - 2.7 Radio license for Europe

165 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210115.2 ConnectingThe voltage supply and signal output a

Page 9

175 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210118. Connect the screen to the internal ground termin

Page 10 - 3 Product description

185 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011+Bus2312( )(-)1Fig. 6: Connection compartment, doub

Page 11 - 3.2 Principle of operation

195 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011+Bus212( )(-)1Fig. 9: Connection compartment, doubl

Page 12

2Quick startVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Quick startThe quick start enables a quick setup with many applications.

Page 13

205 Connecting to the bus systemVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011• Internal check of the electronics• Indication of t

Page 14 - 4 Mounting

216 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210116 Set up with the indicating an

Page 15 - 5.1 Preparing the connection

226 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210111 2Fig. 13: Insertion of the ind

Page 16 - 5.2 Connecting

236 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011 – Edit parameter – Save value•

Page 17

246 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011You can nd in the operating ins

Page 18 - SIM-Card

256 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011100%0%0,5 m(19.68")5 m(196.

Page 19

266 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210114. Enter the suitable distance

Page 20

276 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The "Echo curve" shows

Page 21

286 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The following circumstances caus

Page 22 - 6.2 Adjustment system

296 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011If a false signal suppression ha

Page 23 - 6.3 Parameter adjustment

3Quick startVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The quick start is then nished. For further information see chapter &qu

Page 24

306 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011With radar sensors with standpip

Page 25

316 Set up with the indicating and adjustment moduleVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Menu section Menu item Default v

Page 26

327 Setup with PACTwareVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210117 Setup with PACTware7.1 Connect the PC312Fig. 16: Connecti

Page 27

337 Setup with PACTwareVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Fig. 17: Example of a DTM viewAll device DTMs are available a

Page 28

348 Set up with other systemsVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210118 Set up with other systems8.1 DD adjustment programs

Page 29

359 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210119 Diagnosis, Asset Management and servic

Page 30

369 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Further echo curves: Up to 10 echo curves

Page 31

379 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The following table shows the codes and t

Page 32 - 7 Setup with PACTware

389 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011CodeText mes-sageCause Rectication PA De

Page 33

399 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The following table shows the error codes

Page 34 - 8 Set up with other systems

4ContentsVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Contents1 About this document1.1 Function ...

Page 35 - 9.1 Maintenance

409 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The below tables show typical examples of

Page 36 - 4 Maintenance - blue

419 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210119.4 MeasurementerrorduringllingFault

Page 37

429 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011In Ex applications only one instrument an

Page 38

439 Diagnosis, Asset Management and serviceVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011• Please contact for the return shipment

Page 39 - 9.4 Rectify faults

4410 DismountingVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-12101110 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be

Page 40

4511 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-12101111 Supplement11.1 Technical dataGeneral data316L corresponds to 1.44

Page 41

4611 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210113421Fig. 26: Data of the input variable1 Reference plane2 Measured

Page 42 - 9.6 Software update

4711 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011 Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)Deviat

Page 43

4811 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Gas phase Temperature Pressure1 bar (14.5 psig)10 bar (145 psig)50 ba

Page 44 - 10 Dismounting

4911 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Vessel pressure relating to the ange nominal stagesee supplementary

Page 45 - 11 Supplement

5ContentsVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210119.3 Asset Management function ...

Page 46

5011 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Adjustment elements 4 keysProtection rating Ʋ unassembled IP 20 Ʋ mou

Page 47

5111 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011For that reason the associated approval documents of these instrument

Page 48

5211 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011General Information LAS (Link Active Scheduler) Ye sMaster Capable Ye

Page 49

5311 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Fig. 30: Schematic presentation function block Discret Input (DI)Func

Page 50

5411 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Fig. 32: Schematic presentation function block Output SplitterFunctio

Page 51

5511 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Fig. 34: Schematic presentation function block IntegratorFunction blo

Page 52 - 11.2 Function blocks

5611 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011The following algorithms are available:• Flow compensation, linear• F

Page 53

5711 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011FF desciptor Description UnitSECONDARY_VALUE_1_TYPESecondary value 1

Page 54

5811 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011FF desciptor Description UnitFALSE_SIGNAL_CMD_CON-FIGURATION2Gradient

Page 55

5911 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011FF desciptor Description UnitECHO_CURVE_PLICSCOM_LIMITSParameters as

Page 56 - 11.2 Parameter list

61 About this documentVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210111 About this document1.1 FunctionThis operating instructions

Page 57 - 36510-EN-121011

6011 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Aluminium housing21ø 86 mm (3.39")~ 116 mm (4.57")116 mm (4

Page 58

6111 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011Stainless steel housing~ 69 mm(2.72")ø 79 mm(3.11")117 mm (

Page 59 - Plastic housing

6211 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011VEGAPULS 62, standpipe version with threadø 21,3 mm(0.84")ø 21,3

Page 60

6311 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011VEGAPULS62,standpipeversionwithangeø 39,5 mm(1.56")ø 21,3

Page 61

6411 SupplementVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-12101111.4 Industrial property rightsVEGA product lines are global protect

Page 62 - (1.26")

65INDEXVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011INDEXAAccessories – External indicating and adjustment unit 12 – External ra

Page 63

66NotesVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011

Page 64 - 11.5 Trademark

67NotesVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-121011

Page 65

Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany36510-EN-121011All statements concerning scope of delivery, application, practi

Page 66

72 For your safetyVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210112 For your safety2.1 Authorised personnelAll operations describe

Page 67

82 For your safetyVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210112.5 CE conformityThe device fullls the legal requirements of the

Page 68 - Printing date:

92 For your safetyVEGAPULS 62 standpipe version • Foundation Fieldbus36510-EN-1210112.9 Environmental instructionsProtection of the environment is on