Operating InstructionsTDR sensor for continuous level measurement of bulk solidsVEGAFLEX 82Foundation FieldbusDocument ID: 44221
103 Product descriptionVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309103 Product description3.1 CongurationThe nameplate contains the most importan
Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany44221-EN-130910All statements concerning scope of delivery, application, practi
113 Product descriptionVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910This operating instructions manual applies to the following instrument versions
123 Product descriptionVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309103.3 Packaging, transport and storageYour instrument was protected by packaging
133 Product descriptionVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910For sensors with double chamber housing the interface adapter "DISADAPT&qu
143 Product descriptionVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910If you mount the VEGAFLEX 82 in a bypass tube or standpipe, you have to avoid c
154 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309104 Mounting4.1 General instructionsOn instruments with process tting thread, the hexagon
164 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910vessels, the distance to the vessel wall should be at least 500 mm (19.7 in).During operat
174 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101 2Fig. 6: Installation in non-metallic vessel1 Flange2 Metal sheetConcrete vesselWhen in
184 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Higher sockets or sockets with a bigger diameter can generally be used. They can, however,
194 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig.10:MountingofthesensorwithinowingmediumThe reference plane for the measuring
2Quick startVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Quick startThe quick start procedure enables a quick setup of the instrument with many app
204 MountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101212Fig. 11: Fasten the probe1 Probe2 Retaining sleeveIn case of dicult installation con
215 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309105 Connecting to power supply5.1 Preparing the connectionAlways keep in
225 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101. Unscrew the housing cover2. If a display and adjustment module is i
235 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910the terminal opening is freed. When the screwdriver is released, the ter
245 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091050101+678Bus2312( )(-)11Fig. 15: Electronics compartment, double chambe
255 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309105.5 Wiring plan, Ex-d-ia double chamber housing50101+678Bus2312( )(-)4
265 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910312Fig. 20: View to the electronics compartment1 DIS-ADAPT2 Internal pl
275 Connecting to power supplyVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309105.7 Supplementary electronicsThe radio module PLICSMOBILE is an externa
286 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309106 Set up with the display and adjustment module6.1
296 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101 2Fig. 25: Insertion of the display and adjustment m
3Quick startVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101. In this menu item you can select the application.2. In the menu item "Medium - D
306 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910 – Conrm selected menu – Edit parameter – Save value
316 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910You can nd "Extended adjustment" in the ne
326 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309106.4 Parameter adjustment - Extended adjustmentFor te
336 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910In this menu item you can select which type of medium
346 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Enter the appropriate distance value in m (correspond
356 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Enter the actual distance from the sensor to the prod
366 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Warning:If a linearization curve is selected, the mea
376 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910In the menu item "Lock/unlock adjustment",
386 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910The default setting for the displayed value 1 is &quo
396 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910is displayed as percentage value. The higher the valu
4Quick startVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101. In the menu "Additional settings", menu item "Damping" you can adj
406 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910In this menu item you can simulate measured values vi
416 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Entering a 4-digit PIN protects the sensor data again
426 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Menu item Default value ModiedvalueProbe length Len
436 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Menu item Default value ModiedvalueDevice memory -
446 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Note:Before the data are stored in the sensor, a chec
456 Set up with the display and adjustment moduleVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910In this menu item, the features of the sensor such as
467 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309107 Setup with PACTware7.1 Connect the PC312Fig. 28: Connection of the PC direc
477 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 29: Example of a DTM viewAll device DTMs are available as a free-of-charge
487 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 30: Select quick setup1 Quick setup2 Extended adjustment3 MaintenanceQui
497 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Serial numberHere, you can nd the serial number of your instrument. You cannot
5ContentsVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Contents1 About this document1.1 Function ...
507 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Application - Level measurementThe level measurement refers to the product surf
517 Setup with PACTwareVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910When the setup of the instrument is nished, additional settings can be made. T
528 Set up with other systemsVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309108 Set up with other systems8.1 DD adjustment programsDevice description
539 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309109 Diagnostics and service9.1 MaintenanceIf the device is used correctly, n
549 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910• Display and adjustment moduleFurther echo curves: Up to 10 echo curves ca
559 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910still valid. Plan in maintenance for the instrument because a failure is e
569 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910CodeText mes-sageCause Rectication PA DevS-pecDiagnosisF125Unper-missible
579 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910CodeText mes-sageCause RecticationS600Unpermissi-ble electronics temperatu
589 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910CodeText mes-sageCause RecticationM505No echo avail-able – Level echo can
599 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910• Wherever the sensor displays a constant value, the reason could also be t
6ContentsVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309109.2 Diagnosis memory ...
609 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fault description Error pattern Cause Rectication5. Measured value jumps s
619 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091012Leveltime0Fig. 41: The broken line 1 shows the real level, the continuous
629 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fault description Error pattern Cause Rectication4. Measured value re-main
639 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910In Ex applications only one instrument and one electronics module with resp
649 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 50: Exchange cable or rodInformation:Please keep the stated torque so
659 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309107. Cable: Fasten the cable with three pins, torque 20 Nm (14.75 lbf in)Cab
669 Diagnostics and serviceVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910You can nd detailed information on www.vega.com/downloads and "Approv
6710 DismountingVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091010 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous
6811 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091011 Supplement11.1 Technical dataGeneral data316L corresponds to 1.4404 or 1.4435Materi
6911 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910 Ʋ Instrument weight (depending on process tting)approx. 0.8 … 8 kg (0.176 … 17.64 lbs
71 About this documentVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309101 About this document1.1 FunctionThis operating instructions manual provides al
7011 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091012AB123456345661210842036542100510 15 20 25 30 35 0510 15 20 25 30 35Fig. 52: Max. tens
7111 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091012AB123456345612181614108642069875432100510 15 20 25 30 35 0510 15 20 25 30 35Fig. 54:
7211 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910 Ʋ Cable: ø 6 mm (0.236 in) - PA coated 8 Nm (5.9 lbf ft) Ʋ Cable: ø 6 mm (0.236 in) 20
7311 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309104531245312Fig. 56: Measuring ranges - VEGAFLEX 821 Reference plane2 Probe length L3
7411 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910 Ʋ EEx-d instrument 14 … 32 V DCPower supply by/max. number of sensors Ʋ Fieldbus max.
7511 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091002mm(0.079")-2mm(-0.079")15mm(0.591")10mm(0.394")0,08 m(3.15&qu
7611 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Characteristics and performance dataMeasuring cycle time < 500 msStep response time3
7711 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910120°C(32°F)-40°C(-104°F)55°C(131°F)80°C(176°F)200°C(392°F)150°C(302°F)34-20°C(-68°F)Fig
7811 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910120°C(32°F)-40°C(-104°F)65°C(149°F)80°C(176°F)200°C(392°F)170°C(338°F)34-20°C(-68°F)Fig
7911 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Electromechanical data - version IP 66/IP 68 (1 bar)Options of the cable entry Ʋ Cable
82 For your safetyVEGAFLEX 82 • Foundation Fieldbus44221-EN-1309102 For your safety2.1 Authorised personnelAll operations described in this operatin
8011 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910 Ʋ Non-Ex instrument 9 … 32 V DC Ʋ EEx-ia instrument - Power supply FISCO model9 … 17.5
8111 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Electricial Characteristics Physicial Layer Type Low-power signaling, bus-pow-ered, FIS
8211 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910m(d)% Lin%PrimaryValueSecondaryValue 2SecondaryValue 1TBSensor_Valuemin/maxadjustmentLi
8311 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Function block PID ControlThe function block "PID Control " is a key componen
8411 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 67: Schematic presentation function block Output SplitterFunction block Signal Cha
8511 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 68: Schematic presentation function block Signal CharacterizerFunction block Integ
8611 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 69: Schematic presentation function block IntegratorFunction block Input SelectorT
8711 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 70: Schematic presentation function block Input SelectorFunction block ArithmeticT
8811 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Fig. 71: Schematic presentation function block ArithmeticParameter listThe following ta
8911 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910FF desciptor Description UnitDIAG_MASK_2DIAG_OUT_2DEVICE_IDENTIFICATION Manufacturer ID
92 For your safetyVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Electromagnetic compatibilityInstruments with plastic housing as well as in four-wir
9011 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910FF desciptor Description UnitGU_ID_ENDECHO_CURVE_READ Echo curve dataECHO_EVALUATOR Ech
9111 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910FF desciptor Description UnitNOISE_DETECTION_INFO Increase of the system noiseNOISE_DET
9211 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Aluminium housing in protection rating IP 66/IP 68 (1 bar)120 mm(4.72")~ 105 mm(4.
9311 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Stainless steel housing in protection rating IP 66/IP 68, 1 bar117 mm(4.61")~ 103
9411 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910VEGAFLEX 82, cable version ø 4 mm (0.157 in)/ø 6 mm (0.236 in) - PA coatedL46mm(1.81&q
9511 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910VEGAFLEX 82, cable version ø 6 mm (0.236 in)/ø 11 mm (0.433 in) - PA coated80mm(3.15&q
9611 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910VEGAFLEX 82, rod version ø 16 mm (0.63 in)80mm(3.15")L30mm(1.18")ø 16mm(0.
9711 SupplementVEGAFLEX 82 • Foundation Fieldbus44221-EN-13091011.4 Industrial property rightsVEGA product lines are global protected by industrial p
98INDEXVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910INDEXAAdjustment – Max. adjustment 33 – Min. adjustment 34Adjustment system 30Application 3
99INDEXVEGAFLEX 82 • Foundation Fieldbus44221-EN-130910Service – Hotline 62Setup – Quick start 2Simulation 40Special parameters 44Statusmeldungen