Operating InstructionsVEGAVIB 62- Relay (DPDT)Document ID:29271Vibration
4 Mounting4.1 General instructionsMake sure that all parts of the instrument exposed to the process, inparticular the sensor element, process seal and
The max. permissible pressure is specified in chapter "Technical data"or on the type label of the sensor.The vibrating level switch is a meas
dd10123Fig. 4: Filling in the centre, emptying laterally1 VEGAVIB 622 Discharge opening3 Filling openingThe vibrating element should protrude into the
Fig. 5: Inflowing mediumIn applications such as grit chambers or settling basins for coarsesediments, the vibrating element must be protected against d
5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o
5 Lift the opening levers of the terminals with a screwdriver (seefollowing illustration)Fig. 7: Connection steps 5 and 66 Insert the wire ends into t
14442 3Fig. 8: Material versions, single chamber housing1 Plastic (not with EEx d)2 Aluminium3 Stainless steel (not with EEx d)4 Filter element for pr
6 Set up6.1 General informationThe figures in brackets refer to the following illustrations.On the electronics module you will find the following indica
As a default setting, the potentiometer of VEGAVIB 62 is set to thecomplete right position (> 0.3 g/cm³/0.011 lbs/in³). In very light solidsyou hav
Level Switching status Control lampMode min.Dry run protection53 4(8)(6) (7)Relay deenergized RedFailure of the sup-ply voltage(min./max. mode)any53 4
Contents1 About this document1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . .
7 Maintenance and fault rectification7.1 MaintenanceIf the instrument is used properly, no special maintenance is requiredin normal operation.7.2 Remov
Error Cause RectificationWrong mode selected Set the correct mode on the mode switch (overflowprotection, dry run protection). Wiring should be carriedo
6 Pull out the old electronics module7 Compare the new electronics module with the old one. The typelabel of the electronics module must correspond to
8 Dismounting8.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper
9 Supplement9.1 Technical dataGeneral dataMaterial 316L corresponds to 1.4404 or 1.4435Materials, wetted parts- Process fitting - thread 316L- Process
Sensor length (L) - suspension cable FEP(-40 … +150 °C/-40 … +302 °F)0.6 … 80 m (1.969 … 262.47 ft)Output variableOutput Relay output (DPDT), 2 floatin
Granular size no limitation1)Electromechanical dataCable entry/plug (dependent on the version)- Single chamber housing l 1 x cable entry M20 x 1.5 (ca
9.2 DimensionsVEGAVIB 62 - housing3214~ 69 mm(2.72")ø 79 mm(3.11")117 mm (4.61")M20x1,5/½ NPT~ 59 mm(2.32")ø 80 mm(3.15")112
49 mm(1 59/64")25 mm(63/64")20 mm(25/32")L185 mm (7 9/32")125 mm (4 59/64")ø16 mm (5/8")G1A G1 1/2A ø 29 mm (1 9/64&quo
47 mm(1 27/32")25 mm(63/64")20 mm(25/32")L176 mm (6 59/64")125 mm (4 59/64")ø16 mm (5/8")G1A G1 1/2A ø 29 mm (1 9/64&qu
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Supplementary documentationInformation:Supplementary documents appropria
9.3 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only i
VEGAVIB 62 • - Relay (DPDT) 319 Supplement29271-EN-120418
VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9
1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as
2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist
2.5 Safety label on the instrumentThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fulfills the
3 Product description3.1 StructureThe scope of delivery encompasses:l VEGAVIB 62 point level switchl Documentation- this operating instructions manual
l SIL identification (with SIL rating ex works)With the serial number, you can access the delivery data of theinstrument viawww.vega.com, "VEGA To
3.3 OperationWith the factory setting, products with a density of > 0.05 g/cm³(0.002 lbs/in³) can be measured. It is possible to adapt the instrume